PRO 2Sensor (IHT) Oxygen Sensor Specification
Sensor Type
Zirconium Dioxide Ceramic
Range
0.1% - 20.9%
Accuracy
+/- 5% of reading (Test gases traceable to NPL standards)
IP Rating
IP20 (sensor) IP65 (transmitter)
Housing
Stainless Steel (sensor) Aluminium (transmitter)
Weight
1.2Kg
Connection
6 way plus earth Amphenol Ecomate series
Dimensions
320 x 62mm (max)
Heater Supply
nom. 1.4A @13.1V DC
Output
4-20 mA (0.1% - 20.9% O2) Galvanically isolated Polarity protected
Supply
7-35V DC (Load resistance ≤(Vsupply-7.2)/0.023(Ω))
Isolation
1500V AC for 60 Sec’s
Air Humidity
0-90% RH (non condensing)
Flue Gas Temperature
500 degrees C sustained; 600 degrees C max (200 hour cumulative)
Ambient Temperature
-40 to 85 degrees C
EMC
Emissions EN50081-1 Immunity EN50082-2
Safety
EN61010-1
Temperature Probe
4-20 mA (0-500 degrees C)
Power Supply
100 – 240V AC @ 50/60Hz  40VA nominal